A fast-response and sensitive amperometric biosensor for choline and acetylcholine based on choline oxidase and acetylcholinesterase immobilized on a platinum electrode by co-crosslinking with bovine serum albumin and glutaraldehyde has been developed. The immobilization procedure proved fast and simple, not requiring dedicated manipulation. Detection limits mere in the sub-micromolar range and linearity extended over three concentration decades. After one month of discontinuous use, 70% of the initial sensitivity was still observed. The response time was around 1 s for both analytes making the developed biosensor a suitable detector for flow injection systems and high-performance liquid chromatography.
AMPEROMETRIC SENSOR FOR CHOLINE AND ACETYLCHOLINE BASED ON A PLATINUM-ELECTRODE MODIFIED BY A CO-CROSS-LINKED BIENZYMIC SYSTEM
DE BENEDETTO, Giuseppe, Egidio;
1995-01-01
Abstract
A fast-response and sensitive amperometric biosensor for choline and acetylcholine based on choline oxidase and acetylcholinesterase immobilized on a platinum electrode by co-crosslinking with bovine serum albumin and glutaraldehyde has been developed. The immobilization procedure proved fast and simple, not requiring dedicated manipulation. Detection limits mere in the sub-micromolar range and linearity extended over three concentration decades. After one month of discontinuous use, 70% of the initial sensitivity was still observed. The response time was around 1 s for both analytes making the developed biosensor a suitable detector for flow injection systems and high-performance liquid chromatography.I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.