Long-term bioelectronic implants require stable, hermetic encapsulation. Water and ion ingress are challenging to quantify, especially in miniaturized microsystems and over time. We propose a wireless and battery-free flexible platform leveraging backscatter communication and magnesium (Mg)-based microsensors. Water permeation through the encapsulation induces corrosion of the Mg resistive sensor thereby shifting the oscillation frequency of the sensing circuit. Experimental in vitro and in-tissue characterization provides information on the operation of the platform and demonstrates the robustness and accuracy of this promising method, revealing its significance for in-situ real-time monitoring of implanted bioelectronics.
Wireless, battery-free, and real-time monitoring of water permeation across thin-film encapsulation
Mariello, Massimo;
2024-01-01
Abstract
Long-term bioelectronic implants require stable, hermetic encapsulation. Water and ion ingress are challenging to quantify, especially in miniaturized microsystems and over time. We propose a wireless and battery-free flexible platform leveraging backscatter communication and magnesium (Mg)-based microsensors. Water permeation through the encapsulation induces corrosion of the Mg resistive sensor thereby shifting the oscillation frequency of the sensing circuit. Experimental in vitro and in-tissue characterization provides information on the operation of the platform and demonstrates the robustness and accuracy of this promising method, revealing its significance for in-situ real-time monitoring of implanted bioelectronics.| File | Dimensione | Formato | |
|---|---|---|---|
|
Mariello_NatureComm_2024.pdf
accesso aperto
Descrizione: Articolo
Licenza:
Creative commons
Dimensione
8.14 MB
Formato
Adobe PDF
|
8.14 MB | Adobe PDF | Visualizza/Apri |
I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.